Two-point DAC calibration can correct a linear gain and offset error, but agreement at the two calibration points is not a complete accuracy result. The corrected output still has quantization, nonlinearity, measurement uncertainty, and operating-condition dependence. Validate the residual across the useful range before assigning the calibrated channel an accuracy specification.

Choose valid calibration points

The Analog Devices article on DAC gain and offset calibration describes correcting the transfer relationship using measured errors. Calibration cannot create output range beyond the physical DAC and amplifier limits.

Select two separated points within the linear operating region. Avoid using a rail-limited endpoint as if it were an ordinary gain measurement. Record reference voltage, output load, temperature, and the instrument used to measure the output.

Fit the measured relationship

For measured output y and applied code x, calculate the slope from the difference between the two output readings divided by the code difference. Calculate the intercept from either point. Use the inverse relationship to determine the code for a desired output.

Keep units and code conventions explicit. A voltage in millivolts combined with coefficients stored in volts can produce a plausible-looking but incorrect correction factor.

Check code headroom and rounding

Evaluate the corrected code required at the lowest and highest requested outputs. If the result lies outside the available code range, the requested calibrated span must change or the hardware must provide more headroom.

For example, correcting a positive gain error may reduce the code needed near the top of the range, while correcting an offset can consume code margin at one end. Check both effects together rather than clipping the result silently.

Measure independent verification points

Test several points that were not used in the fit, including relevant region boundaries. Plot requested output, measured output, and residual error. A curved residual reveals behavior that a straight-line correction cannot remove.

Repeat important points at changed load and temperature. If calibration is intended to apply across those conditions, the verification must cover them or the remaining error must be budgeted.

Preserve uncalibrated and calibrated data so the improvement is visible and the correction can be audited. Store coefficients with channel identity, units, calibration conditions, and the software interpretation.

The release evidence should state the verified span and maximum observed residual with measurement uncertainty considered. Passing the two fitted points demonstrates that the arithmetic reproduces its inputs; independent points establish whether the corrected channel is useful throughout the range the product promises.