A current transformer with ample high frequency bandwidth may still misrepresent a long current pulse. Peak current, current time product, and low frequency response are separate constraints. Review all of them before interpreting a falling pulse top as a change in the PCB load.
Identify the probe technology
This review applies to AC current transformers, not automatically to Hall effect or other DC capable probes. The Tektronix CT1, CT2, and CT6 data sheet identifies these devices as non-DC current transformers and specifies peak pulse current, ampere second limits, droop behavior, and termination requirements. Use the limits for the exact probe rather than transferring a rating between models.
Write down pulse amplitude, width, repetition pattern, polarity, and any DC component. Include startup and abnormal operating pulses. A short repetitive pulse and a sustained overload can have the same peak amplitude but place very different demands on the measurement system.
Calculate pulse area
For a rectangular pulse, current multiplied by duration gives the ampere second product. A hypothetical 2 A pulse lasting 10 microseconds has an area of 20 ampere microseconds. For a varying pulse, use its current time integral according to the manufacturer’s definition and conditions.
Compare both peak current and pulse area with the probe specification. Passing one check does not imply passing the other. Also assess continuous heating and repetition limits. Do not assume an off interval always restores the probe to an identical magnetic state without reviewing the specified operating conditions.
Recognize droop separately
A transformer has a finite low frequency response, so a flat current pulse can produce a measured output that declines during its duration. That behavior is distinct from exceeding a maximum rating. Compare the pulse width with the documented response and decide whether the expected droop is acceptable for the measurement objective.
If the purpose is to determine steady load current, select a suitable DC measurement method. If the purpose is to inspect a fast current edge, the transformer may remain useful even when it cannot establish the sustained current level. State that distinction in the result.
Preserve the specified signal path
Use the required termination and cable arrangement. An incorrect oscilloscope input impedance changes sensitivity and response. Record conductor routing through the aperture, the current direction, probe model, termination, and waveform timing.
Compare a suspicious long pulse with a suitable shunt or DC capable probe when practical. Agreement near the leading edge but disagreement later is a reason to examine low frequency response. It is not sufficient evidence that the circuit itself contains a decaying current.